Balance Setup; Balance Set-Up Sub-Menu; Stable Range; Filter Level - OHAUS Explorer EX1241 Instruction Manual

Hide thumbs Also See for Explorer EX1241:
Table of Contents

Advertisement

EXPLORER BALANCES
5.5

Balance Setup

Enter this sub-menu to customize Balance display functionality.
5.5.1

Balance Set-up sub-menu

Stability
Indicator Range
Gross Indicator
Factory default settings are shown below in bold.

Filter Level

Legal For Trade
5.5.2

Stable Range

Set the amount the reading can vary while the stability symbol remains on.
0.5 D
= 0.5 graduations
1 D
= 1 graduation
2 D
= 2 graduations
3 D
= 3 graduations
5 D
= 5 graduations
Note: When LEGAL FOR TRADE (below) is set to ON, the setting is forced to 1 D.
The setting is locked when the Security Switch is set to the ON position.
5.5.3
Filter level
Set the amount of signal filtering.
LOW
= faster stabilization time with less stability.
MEDIUM
= normal stabilization time with normal stability.
HIGH
= slower stabilization time with more stability.
The setting is locked at the current setting when the security switch is set to
the ON position.
5.5.4

Auto Zero Tracking

Set the automatic zero tracking functionality.
OFF
= disabled.
0.5 D
= display maintains zero up to a drift of 0.5 graduation
1 D
= display maintains zero up to a drift of 1 graduation per
3 D
= display maintains zero up to a drift of 3 graduations per
Note: When Legal for Trade is set to ON, the AZT setting is forced to 0.5 D.
The OFF setting is still available. The setting is locked at the current setting when the
security switch is set to the ON position
Auto Zero
Tracking
Graduations
per second
second.
second.
Auto Tare
Date & Time
EN-97