4. Calibration masses value will be shown in the screen.
After the display shows "Place weights" and "50.0000 g",
place weight(s) of 50 g on the pan for calibration.
5. Remove the weight(s) of 50 g from the pan. After a while,
"100.0000 g" will be displayed on the screen. Please
place weight(s) of 100 g on the pan for calibration.
6. Once the linearity calibration is completed successfully,
"Linearity done" will display.
Press any button to return to the previous screen.

5.3 Balance Setup

Enter this sub-menu to customize balance functionality.
Note: The factory default settings are shown below in bold.

5.3.1 Language

Set the language displayed for menus and displayed
messages.
English
Chinese
Japanese
Korean

5.3.2 Filter Level

Set the amount of signal filtering.
Low
= faster stabilization time with less stability.
Medium
= normal stabilization time with normal stability.
High
= slower stabilization time with more stability.

5.3.3 AZT (Auto Zero Tracking)

Set the automatic zero tracking functionality.
Off = disabled.
0.5 d = display maintains zero up to a drift of 0.5 graduation per second.
1 d = display maintains zero up to a drift of 1 graduation per second.
3 d = display maintains zero up to a drift of 3 graduations per second.