Lauda V1R20 Operation Manual page 28

Command touch operating unit
Table of Contents

Advertisement

Fig. 22: List of programs in the programmer
Types of segment
Creating a program
28 / 51
1.
Open the Home window.
2.
Touch the [Programs] tile.
The list of available programs appears.
3.
Use the buttons under the list of programs to operate this:
[Edit]: edit program or view without change.
[Delete]: remove program from list.
[Undo delete]: reverse last delete operation.
[Copy]: copy the selected program to the next line.
[New]: create a new program without segments at the end of the
list.
The [Copy] and [New] buttons are only available if the list
has less than 100 entries.
Using the graphical display in the temperature profile, the following types of
segment can generally be distinguished:
Ramp; described by the following values:
Time limit for the duration from the beginning to the end of the
segment.
Target temperature which should be achieved at the end of the
segment.
Jump (temperature jump); corresponds to a ramp without a time limit.
With the time limit value = 0, the constant temperature equipment tries
to reach the target temperature as quickly as is technically possible.
Maintain (temperature maintaining phase); corresponds to a ramp
without a temperature change. The value of the previous segment is
entered as the target temperature. Thus, the constant temperature
equipment keeps this temperature constant over the entire segment
duration (time limit).
Pump level 0 ; can be inserted anywhere in a program to terminate it
prematurely. When the program starts, a message is displayed informing
you of the existence of the Pump level 0 segment. When this segment
is reached, the constant temperature equipment switches to Standby
mode.
In the Edit program menu, you can edit the settings of the selected program
and create the segments for the required program sequence:
1.
First carry out the following settings:
Name - Enter a name with which the program will be displayed in
the program list and other menus.
Cycles - Specify how often the program should be run in succession
by entering a number.
Command Touch operating unit
V1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lrt 923

Table of Contents