Please read these operating instructions before unpacking and putting the unit into operation. Follow the instructions precisely as described herein. The instruction manuals on our website www.kobold.com are always for currently manufactured version of our products. Due to technical changes, the instruction manuals available online may not always correspond to the product version you have purchased.
In case of damage, please inform your parcel service / forwarding agent immediately, since they are responsible for damages during transit. Scope of delivery: The standard delivery includes: Pressure Transmitter model: SEN-98/-99 4. Product Label SEN-98/-99 K03/0322 page 5...
The pressure transmitter turns the input pressure into an output electrical signal. The electrical signal changes in proportion to the input pressure level. Any use of the Pressure Transmitter, model: SEN-98/-99, which exceeds the manufacturer’s specification may invalidate its warranty. Therefore, any resulting damage is not the responsibility of the manufacturer.
Figure 2 – Wires connection Warning! The IP grade according to standard EN-60529-1:1992 is guaranteed only if the female connector equipped with a connection cable is mounted on the instrument and all the other components are assembled correctly. page 8 SEN-98/-99 K03/0322...
In models the zero and the full scale range can be adjusted as follows (see figure 3): - Connector disassembling (see fig. 1) - Zero adjustment (Z) - Span adjustment(S) Figure 3 - Adjustment In case a complete recalibration is necessary please contact KOBOLD. SEN-98/-99 K03/0322 page 9...
≤230 psi (≤ 16 bar) O-ring: Electric connection: *EN 175301-803 form A, exit for cables ø 0.23 ... 0.35” (6 ... 9 mm) Protection degree: IP 65 as per IEC 529 / EN 60529 Weight: 0.39 Ibs (0.18 kg) page 10 SEN-98/-99 K03/0322...
Page 11
-30...100 in Hg/psi -30...150 in Hg/psi 0...15/0...30 0...60 0...100/0...160 0...300 0...600 1450 1740 0...1000/0...1500 2900 3625 0...2000/0...3000 7250 8700 0...5000/0...6000 8700 11600 0...10000 11600 13050 Other ranges available on demand. Units of measurement available in MPa, kPa too SEN-98/-99 K03/0322 page 11...
B 055 = 0...4 bar 175301-803) B 065 = 0...6 bar B 075 = 0...10 bar SEN-9830 B 085 = 0...16 bar (M12x1 male B 095 = 0...25 bar electrical connection) 1)3) SEN-9840 (2 m PVC cable, page 12 SEN-98/-99 K03/0322...
10…30 8…30 14…30 Ground (please refer to chapter 6) 11. Dimensions (in mm) L (mm) A – G ½ A (standard) B – G ¼ A F – ½ -14 NPT G – ¼ -18 NPT SEN-98/-99 K03/0322 page 13...
SEN-98/-99 12. EU Declaration of Conformance We, KOBOLD Messring GmbH, Hofheim-Ts, Germany, declare under our sole responsibility that the product: Pressure Transmitter Model: SEN-98/-99 to which this declaration relates is in conformity with the standards noted below: EN 61326-1:2013 Electrical...