Display Example
P:10.0 S
T RR:15
I/E: 1:1.0 (50%)
LK:123 S
S EV:AS
MV:123.4 VT:1234
Max IPAP
15.0 cmH2O
Min EPAP
5.0 cmH2O
Pressure Support
5.0 cmH2O
IPAP
10.0 cmH2O r 5.0
EPAP
5.0 cmH2O r 5.0
Set Pressure
10.0 cmH2O
Delay Time
20 minutes
Delay Pressure
4.0 cmH2O
IPAP Rounding
3 (1234 ms)
EPAP Rounding
3 (1234 ms)
Insp. Trigger
5
Exp. Trigger
5
Tubing Length
1.8 m (6 ft)
SmartFlex Menu
Press 
Enable Menu
Press 
SE-DV57-1
Description
Real time info screen. Top row shows measured pressure (P), Trigger
source (Spontaneous or Timed), breath state (dash in center of top row
goes high during inhale and low during exhale). Bottom row shows I:E
ratio and % inhale.
Real time info screen. Top row shows Leak rate (LK), Trigger source
(Spontaneous or Timed), breath state (dash in center of top row goes
high during inhale and low during exhale), Event Markers (EV), Bottom
row shows Minute Volume (MV) and Tidal Volume (VT)
Maximum IPAP pressure level that will be administered. Range: (Min
EPAP + Pressure Support) - 25.0 cm H
Minimum EPAP pressure level that will be administered. Range: 3.0 cm
H
0 - (max IPAP - Pressure Support), increments of 0.5 cm H
2
Difference between IPAP and EPAP Pressure. Range: 0 - (Max IPAP -
Min EPAP), 12 cm H
This is a prescription setting which controls the pressure during inhale.
This also shows the pressure delta (5.0 in example) between inhale and
exhale pressure settings
This is a prescription setting which controls the pressure during exhale.
This also shows the pressure delta (5.0 in example) between inhale and
exhale pressure settings
This is a prescription setting which controls the operating pressure.
Selectable delay time setting. Your provider has set the delay time.
This is a prescription setting which controls the operating pressure
during the delay time.
This is a setting which controls the comfort rounding when switching to
IPAP pressure. Settings are 0 thru 5 and show measured rise time of
the last breath in milliseconds.
This is a setting which controls the comfort rounding when switching to
EPAP pressure. Settings are 0 thru 5 and show measured rise time of
the last breath in milliseconds.
This is a setting which controls the sensitivity of the inspiratory trigger. It
can be set from 1 to 10 (1 is most sensitive, 10 is least sensitive).
This is a setting which controls the sensitivity of the expiratory trigger. It
can be set from 1 to 10 (1 is most sensitive, 10 is least sensitive).
The standard tubing length supplied with the device is 6 ft. (1.8 m).
However, you can adjust your SleepCube to accommodate an air supply
tubing length of 10 ft. (3.0 m). Replacement tubing can be obtained
through your home care provider.
Pressing the down key enters the SmartFlex menu. Refer to SmartFlex
Menu section for details
Pressing the down key enters the Enable menu. Refer to next section for
Enable Menu details.
O, increments of 0.5 cm H
2
O max, increments of 0.5 cm H
2
AutoBilevel
O.
2
AutoBilevel
O
2
AutoBilevel
O
2
CPAP Mode
CPAP Mode
Mode
All
All
Mode
Mode
Mode
Bilevel S
Modes
Bilevel S
Modes
All
All
All
Bilevel
Modes
All
Bilevel
Modes
All
Bilevel
Modes
All
Bilevel
Modes
All
All
EN - 23