Dynamic Short-Term Pressure Accuracy with and without DV5HH Humidifier (Per
ISO 17510-1:2007)
Waveform
Volume [mL]
Sinusoidal
cycle with I:E =
1:1
Maximum Flow-Rate (per ISO 17510-1:2007)
Measured pressure at the patient
connection port (cmH
O)
2
Average flow at the patient connection port
(l/min)
Long Term Static Pressure Accuracy ................................................................................................................. +/- 0.5 cmH
Additional Specifications
Equipment classification with respect to protection from electric shock ........................................................................ Class II
Degree of protection from electric shock ................................................................................................. Type BF Applied Part
Degree of protection against ingress of liquids ......................................................................................IPXI Drip-proof vertical
Mode of operation .................................................................................................................................................... Continuous
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen or nitrous oxide.
EN - 16
Breath Rate [min-1]
500
10, 15, 20
3.0 cmH
Set Pressure
[cmH
3.0
9.0
14.0
20.0
25.0
Test Pressures
O
9.0 cmH
O
2
2
3.0
8.9
95.0
160.5
Pressure Accuracy
O]
[Pk-Pk cmH
2
14.0 cmH
O
20.0 cmH
2
13.0
19.0
172.2
158.5
O]
2
0.5
0.5
1.0
1.0
1.0
O
25.0 cmH
O
2
2
24.0
139.7
O
2
SE-DV57-1