During the sequence, the LEDs will indicate the actual output
voltage level as affected by load, not the no load output. Refer to the
graph of typical output voltage under various load conditions at the
rear of these instructions.
If none of the proving unit LED's illuminate when performing the
above tests check the condition of the proving unit batteries and
replace them if required (see section 4.1).
3.3 Test Lamp, Voltage Tester /Indicator Cables
During the above tests, emphasis should also be placed upon the
flexing of the UUT cable along its length, and particularly at the
entry points to the hand held elements, to confirm that the cable has
not fractured.
It may be necessary to perform this test a number of times as the
proving unit times cut out after 10s.

3.4 CALCHECK

The CALCHECK feature may be used to check that a multifunction
tester, insulation tester or continuity tester is maintaining ongoing
accuracy between calibrations. There are two insulation test
resistance values, 0.5M
and 1M
Ω. There are also two continuity test
0.5
and 1
Ω.
resistance values,
One probe of the tester being checked should be placed on the
desired resistance value and the other probe should be placed on
the COM terminal, following the instruction manual for the tester
being used.
The CALCHECK feature is not designed to be used to check loop
testers, where the high test current will damage the proving device.
5