FLASH TESTER
Email: sales@martindale-electric.co.uk
web: www.martindale-electric.co.uk
INSTRUCTION MANUAL
LITE3640
ISSUE 7 07/07
Martindale Electric Co Ltd.

Advertisement

loading

Summary of Contents for Martindale Electric Metrohm E3640

  • Page 1 ® Metrohm Martindale Electric Metrohm House, Imperial Park, E3640 Imperial Way, Watford, Hertfordshire, WD24 4PP, UK T: 01923 441717 F: 01923 446900 FLASH TESTER Email: sales@martindale-electric.co.uk web: www.martindale-electric.co.uk INSTRUCTION MANUAL LITE3640 ISSUE 7 07/07 Martindale Electric Co Ltd.
  • Page 2 Martindale Electric Co Ltd reserves the right to discontinue models at any time, or to change specifications, price or design, without notice and without incurring any obligation.
  • Page 3 9-10 To register a claim under the provisions of this warranty, return the ENVIRONMENTAL instrument or test equipment to Martindale Electric Co Ltd., Metrohm House, Penfold Trading Estate, Imperial Way, Watford WD24 4YY, UK. SPARES AND ACCESSORIES Upon our inspection of the product, we will advise you of the outcome of LIMITED WARRANTY your claim.
  • Page 4: Safety Rules

    Bump Test IEC 68-2-29 SAFETY RULES Vibration Test: BS EN 61010 - clause 8.3 The E3640 has been designed with safety in mind. However, no design can Impact Test: BS EN 61010 - clause 8.2 completely protect against incorrect use. Electrical circuits can be dangerous and Drop Test: BS EN 61010 - clause 8.4 even lethal through lack of caution or poor safety practices.
  • Page 5: General

    Trip Level Range: 0.3 to 3mA 50/60Hz 2 DESCRIPTION Setting Accuracy: ±5% of full 2.1 GENERAL Trip Response Leakage: Will trip for fault duration >5mS The E3640 is an AC flash tester for leakage and breakdown testing. It is but not for <100μS designed for voltage testing of electrical apparatus, components and Breakdown: Will trip for fast transients of...
  • Page 6: Controls

    limits are exceeded the output will shut down until the voltage control is 4.APPLICATIONS turned to reset. Illuminated displays show that the tester is switched on and the ‘kV ON’ lamp indicates that high voltage may be present at the outlet. 4.
  • Page 7: Controls

    To clear the displays the output voltage control is returned to reset. 2.2.3 MODE SELECT Three modes of operation are selectable. NOTE: In the burn mode the trip circuit is inoperable and the Breakdown:- The tester will trip should breakdown occur in the tester will continue to drive current into a fault.
  • Page 8: Performing A Test

    2.2.8 HIGH VOLTAGE SOCKET Trip:- In the trip mode, mains frequency currents associated with leakage and short circuits and high frequency This unique safety socket has an interlock which prevents accidental arcing currents that are produced by breakdown are disconnection of the HV gun probe lead. detected.