INTERFACE
Instruments
D-63533 Mainhausen
HM8112-3b
PRECISION MULTIMETER
Made in Germany
Attention!
Voltages exceeding one of the following values are
regarded potentially dangerous or even lethal:
st
1
30 V
rms
nd
2
42.4 V
peak
rd
3
60 V
DC
Connecting higher voltages than those is only allo-
wed to skilled personnel who are familiar with the
dangers incurred. The relevant safety regulations
are to be strictly observed!
V/SENSE (4mm safety sockets)
Connection of measuring cables for:
voltage measurement
frequency measurement
4 wire resistance measurement (SENSE)
continuity test
temperature measurement by a thermocouple
4 wire temperature measurement by a P–temperature
sensor (SENSE)
The maximum voltage between HI and LO case
(ground) must not exceed 850 V-
The maximum voltage between LO and case
(ground) may not exceed 250 V
A/SOURCE (4mm safety socket)
Connection of measuring cables for:
current measurement, max. 1 ampere
2 wire resistance measurement
4 wire resistance measurement (SOURCE)
4 wire temperature measurement by a PT – temperature
sensor (SOURCE)
continuity test up to 10 Ω
The maximum current may amount to 1 A
STOP
STOP
LOW (4mm safety connectors)
Ground connection for inputs
high-impedance DC-isolated.
Fuse in the current measuring circuit
The shunt is fuse-protected. The fuse (FF) is located in a fuse
holder. The measuring circuit is designed for a maximum al-
lowable measurement current of 1 amper
Serien-Nummer
or 600 V
.
peak
DC
!
rms
!
eff
and
. Both connectors are
R e m o t e O p e r a t i o n
115/230 V ±10% / 50-60 Hz
Watts (max.): 8
Fuse: IEC127 - III
115 V: 0.2 A / 230 V: 0.2 A
5 x 20 mm slow blow,
träge, temporisé, lento
SCANNER
Replacement of the fuse is only allowed, after the
instrument was disconnected from the mains!
A repair of a defective fuse or bypassing the fuse is
very dangerous and absolutely prohibited!
STOP
Replacement of the measuring circuit fuse
The measuring circuit fuse
A replacement of the fuse is only allowed, if no voltage is applied
to the measuring connectors! Therefore all terminals V/SENSE
, ground
and A/SOURCE
cover of the fuse holder has to be turned ccw with a screw driver
having a suitable blade. As the cover can be turned it has to be
pushed by the srew driver into the fuse holder. The cover with
the fuse can then be easily taken out. Replace the defective fuse
by a new fuse of the same type having the same trip current. A
repair of a damaged fuse or the use of other means for bypassing
the fuse is very dangerous and absolutely prohibited! Damages
incurred will void the warranty.
Power receptacle with power switch
Power receptacle for connecting the line cord with according
to DIN49457.
Interface
TheRS-232 interface is on the rear panel of the HM8112-3
(9 pin D-sub socket). The bidirectional interface of HM8112-3 can
receive data (commands) from an external device (PC) or send
data (measurement values and parameters).
The following options are available: USB (HO870), IEEE-488
(HO880)
Voltage selector
Choice of mains voltage (115 V / 230 V).

Remote Operation

The serial interface is electrically isolated from the measuring
circuit (Options: USB, IEEE-488).
The instrument is programmable by a PC. Functions and ranges
can be selected and measurement values stored in the instru-
ment can be read out.
The HM8112-3 is connected to another instrument
by a 1:1 interface cable. It is recommended to use
a 9 to 25 pin standard adapter if a PC with a 25 pin
HINT
COM port isconnected.
230 V
VOLTAGE
SELECTOR
115 V
is accessible from the front panel.
should be disconnected. The
Subject to change without notice