8. Error Messages

If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
Error
Description Potential cause and remedy
«ERR 1» Signal too
The pulse signals on the cuff are too weak.
weak
Re-position the cuff and repeat the
measurement.*
«ERR 2» Error signal During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle tension.
Repeat the measurement, keeping your
arm still.
«ERR 3» No pres-
An adequate pressure cannot be gener-
sure in the
ated in the cuff. A leak may have occurred.
cuff
Check that cuff and pump ball are correctly
connected and that the cuff is not fitted too
loosely. Replace the batteries if neces-
sary. Repeat the measurement.
«ERR 5» Abnormal
The measuring signals are inaccurate and
result
no result can therefore be displayed. Read
through the checklist for performing reli-
able measurements and then repeat the
measurement.*
«HI»
Pulse or cuff
The pressure in the cuff is too high (over
pressure
299 mmHg) OR the pulse is too high
too high
(over 200 beats per minute). Relax for
5 minutes and repeat the measurement.*
«LO»
Pulse too
The pulse is too low (less than 40 beats
low
per minute). Repeat the measurement.*
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.

9. Safety, Care, Accuracy Test and Disposal

Safety and protection

 Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
BP N1 Basic
 This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
 This device comprises sensitive components and must be treated
with caution. Observe the storage and operating conditions
described in the «Technical Specifications» section.
 Protect it from:
- water and moisture
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
 The cuffs are sensitive and must be handled with care.
 Do not exchange or use any other kind of cuff for measuring
with this device.
 Only pump up the cuff once fitted.
 Do not use this device close to strong electromagnetic fields such
as mobile telephones or radio installations. Keep a minimum
distance of 3.3 m from such devices when using this device.
 Do not use this device if you think it is damaged or notice
anything unusual.
 Never open this device.
 If the device is not going to be used for a prolonged period the
batteries should be removed.
 Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
 The measurement results given by this device is not a diagnosis.
It is not replacing the need for the consultation of a physician,
especially if not matching the patient's symptoms. Do not rely on
the measurement result only, always consider other potentially
occurring symptoms and the patient's feedback. Calling a doctor
or an ambulance is advised if needed.
EN
5