PCE Instruments PCE-313A Manual

PCE Instruments PCE-313A Manual

Humidity data logger
Hide thumbs Also See for PCE-313A:

Advertisement

Quick Links

Manual
Humidity Data Logger
PCE-313A
Version 3.1
Date of creation: 15.04.2015
Date of last change: 20.05.2019
PCE Americas Inc.
711 Commerce Way
Suite 8
Jupiter
FL-33458
USA
From outside US: +1
Tel: (561) 320-9162
Fax: (561) 320-9176
info@pce-americas.com
www.pce-instruments.com/english
www.pce-instruments.com
PCE Instruments UK Ltd.
Unit 11
Southpoint Business Park
Ensign way
Hampshire / Southampton
United Kingdom, SO31 4RF
From outside UK: +44
Tel: (0) 2380 98703 0
Fax: (0) 2380 98703 9
info@pce-instruments.co.uk

Advertisement

Table of Contents
loading

Summary of Contents for PCE Instruments PCE-313A

  • Page 1 PCE Americas Inc. PCE Instruments UK Ltd. 711 Commerce Way Unit 11 Suite 8 Southpoint Business Park Jupiter Ensign way FL-33458 Hampshire / Southampton United Kingdom, SO31 4RF From outside US: +1 From outside UK: +44 Tel: (561) 320-9162 Tel: (0) 2380 98703 0...
  • Page 2: Table Of Contents

    How to set the sampling rate ....................... 10 5.7.8 How to format the SD card ......................10 Mains adaptor ........................... 11 System reset ..........................11 RS-232 interface ....................11 Patents ........................11 Batteries ......................... 12 Contact ........................12 PCE Instruments UK ........................ 12 PCE Americas .......................... 12...
  • Page 3: Introduction

    SD memory card with a capacity of 1 … 16 GB. The values saved on the SD card as an xls file can be transferred to a PC for further analysis. The PCE-313A can work independently, without the user being present during the measurement as it can be programmed via the keypad in advance.
  • Page 4: Specifications

    Manual Specifications Measurement ranges - temperature 5 ... 95 % RH - relative humidity 0 ... + 50 °C - K-type -50 ... +1300 °C - calculated dew point -25.3 ... +48.9 °C - calculated wet bulb temperature -21.6 ... +50 °C Resolution - temperature 0.1 % RH...
  • Page 5: System Description

    Manual System description Display Power key (ESC, backlight) Hold key (function key, next key) REC key (enter key) SET key ( key, check time) Logger key ( key, check sampling rate) Sensor input socket (humidity and temperature sensor) K- / J-type input SD card slot RS-232 output Reset key...
  • Page 6: Instructions

    Manual Instructions How to select the function 1. Switch on the device by pressing the power key (2). If you hold the power key for more than 2 seconds, this will switch off the instrument. 2. The device has 4 functions: a) humidity and temperature b) dew point c) wet bulb...
  • Page 7: Max / Min Function

    Manual Max / Min function 1. This function saves the maximum and minimum values of a series of measurements. You can activate this function by pressing the REC key (4). “REC” is displayed. 2. When the “REC” icon appears in the display, you can retrieve the maximum value by means of the REC key (4).
  • Page 8: Manual Data Logger Function

    Manual 5.5.3 Manual data logger function a) Set the sampling rate to 0 seconds (see settings chapter). Press the REC key (4) once so that the REC symbol appears in the display. After that, press the logger key (6). The REC symbol will start flashing and you hear the beeper.
  • Page 9: How To Transfer Data To A Pc

    Manual How to transfer data to a PC 1. After saving data to the SD card by means of the meter, remove the SD card from its slot (9). 2. Insert the SD memory card into the card reader of your computer. 3.
  • Page 10: How To Set The Date And Time

    Manual 5.7.1 How to set the date and time 1. When “dAtE” is displayed, confirm your selection by means of the enter key (4). You can now set the value by means of the (5) and the keys (6). The year comes first. After setting the value, press the enter key (4) to get to the next value.
  • Page 11: Mains Adaptor

    Manual Mains adaptor In addition to battery power supply, the device can also be powered by a 9 V dc mains adaptor. To do so, please use the 9 V input socket (12). If you use the device with the mains adaptor, it is always on. Thus, the power key (2) will not work.
  • Page 12: Batteries

    In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use them or give them to a recycling company which disposes of the devices in line with law. If you have any questions, please contact PCE Instruments. Contact If you have any questions about our range of products or measuring instruments please contact PCE Instruments.