Specifications; Electromagnetic Emissions And Immunity; Emc Guidance - Philips DL8760/15 Manual

Table of Contents

Advertisement

Problem

Specifications

Power supply
Display
Measurement method Oscillometric method
Measurement range
Accuracy
Normal operating
condition
Storage and
transportation
conditions
Measurement
perimeter of the
upper arm
Net weight
External dimensions
Accessories
Mode of operation
Degree of protection
Protection against
ingress of water
Device classification
Frequency band
Modulation
Effective radiated
power
Caution: No modification of this equipment is
allowed.
Electromagnetic emissions and
immunity
The device is approved according to EMC safety
standard EN 60601-1-2. It is designed to be used in
typical domestic environments.
Hereby, Philips Consumer Lifestyle BV declares
that the radio equipment type Bluetooth LE is in
compliance with Directive 2014/53/EU.
The full text of the EU declaration of conformity is
available at the following internet
address: www.philips.com/support

EMC Guidance

- The Blood Pressure Monitor needs special
precautions regarding EMC and needs to be
installed and put into service according to the
EMC information provided in the accompanying
documents.
- Wireless communications equipment such as
wireless home network devices, mobile phones,
cordless telephones and their base stations,
walkie-talkies can affect this equipment and
should be kept at least a distance d = 3.3 m
away from the equipment.
Note: As indicated in IEC 60601-1-2:2007 for ME
equipment, a typical cell phone with a maximum
output power of 2 W yields d = 3.3 m at an
immunity level of 3V/m.
Possible
cause
The blood
pressure
monitor and
mobile
device are
more than 5
meters away
from each
other.
You selected
the wrong
profile on the
blood
pressure
monitor.
3.7V 1000mAH built-in
rechargeable li-polymer
battery, 6V 1A DC charger
Display with white
LED backlight
Visible area = 86.1mm (L)
x 24mm (W)
Rated cuff pressure:
0kPa-40kPa
(0mmHg-300mmHg).
Measurement pressure:
5.3kPa-30.7kPa
(40mmHg-230mmHg).
Heart rate: 40-199 beats
per minute
Pressure:
5°C-40°C within ±0.4kPa
(3mmHg) heart rate: ±5%
of measurement result
on display
Temperature: 5°C to
40°C. Relative humidity:
≤85%RH. Atmospheric
pressure: 86kPa to
106kPa
Temperature: -20°C to
60°C. Relative humidity:
10% to 93%. Atmospheric
pressure: 50kPa to
106kPa
About 22cm-42cm
Approx. 265g
Approx.
m
DC charger, user manual
Continuous operation
Type BF applied part
IP22, This means:
protected against access
to hazardous parts with a
finger and against
vertically falling water
drops when tilted up to
15 degrees.
Battery Powered Mode:
Internally Powered ME
Equipment. DC charger
charged mode: Class II
ME Equipment.
2402-2480 MHz
GFSK
Max. +5dBm
Solution
Place your mobile
device closer to the
blood pressure
monitor.
Select the correct
user profile on the
blood pressure
monitor before
your measurement.
Otherwise the data
cannot be
transmitted to your
app. Repeat the
measurement with
the correct profile
selected
130.9mm×73mm×29.4m­

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dl8760

Table of Contents