- Common arrhythmias (such as atrial or
ventricular premature beats or atrial fibrillation)
and peripheral artery disease / arteriosclerosis
can affect the performance (accuracy) of this
blood pressure monitor. Please consult your
doctor how to best use this blood pressure
monitor if you suffer from any of these
conditions.
- Only use this device for its intended purpose as
described in this user manual.
- Do not confuse self-monitoring with self-
diagnosis. This device allows you to monitor
your blood pressure. Do not begin or end
medical treatment based on the measurement
results. Always consult your doctor for
treatment advice.
- Do not take any therapeutic measures on the
basis of a self-measurement. Never change
prescribed medication without consulting your
doctor. Consult your doctor if you have any
questions about your blood pressure.
- If you are taking medication, consult your
physician to determine the most appropriate
time to measure your blood pressure.
- This device is not intended for use on
extremities other than the arm or for functions
other than obtaining a blood pressure
measurement.
- If the cuff pressure exceeds 40kPa (300mmHg),
the unit will deflate automatically. If the cuff
does not deflate when pressures exceeds
40kPa (300mmHg), detach the cuff from the
arm and press the 'on' button to stop inflation.
- Do not attach the cuff on the same arm on
which other monitoring medical electrical
equipment is attached simultaneously, because
this could cause temporary loss of function of
those simultaneously-used monitoring medical
electrical equipment.
- Never attach the cuff on injured skin, an injured
arm or an arm under medical treatment as this
can cause further injury.
- Do not use the device in case of existing
polyester or nylon material allergies.
- This device is not washable. Never immerse the
device in water and do not rinse it under the
tap.
- This device is not suitable for continuous
monitoring during medical emergencies or
operations.
- This device cannot be used with HF (High
Frequency) surgical equipment at the same
time.
- Never use compressed air, scouring pads,
abrasive cleaning agents or aggressive liquids
such as petrol or acetone to clean the device.
- Do not use the adapter in or near wall sockets
that contain or have contained an electric air
freshener to prevent irreparable damage to the
adapter.
- If the battery can no longer be recharged or
when the device does not function properly
anymore (see 'Specifications'), please contact
the Philips Consumer Care center in your
country.
- Keep the device away from fire and heat
sources, as the battery can overheat, causing
fire or bursting. The battery could explode
causing injury or death.
- After charging, remove the small plug from the
device and remove the adapter from the wall
socket.
- The equipment is not AP/APG equipment and
is not suitable for use in the presence of a
flammable anesthetic mixture with air, with
oxygen or nitrous oxide.
- To avoid measurement errors, do not use the
device near strong electrical or magnetic fields,
for example magnets, radio transmitters,
microwave ovens.
- To avoid measurement errors, do not use the
device near a strong electromagnetic field
radiated interference signal or electrical fast
transient/burst signal.
- Use this device under the right environmental
conditions as indicated in this user manual. If
not, this could affect the performance, lifetime
of the device and measurement results.
- Use a soft cloth to clean the whole unit. Do not
use any abrasive or volatile cleaners.
- Only use the DC charger supplied (model: AC
power adapter KH0601000EW) to charge the
device.
- The device does not need to be calibrated for
two years of normal use. After two years the
measurements may become less accurate.
- If you have any problems with this device, such
as setting up, malfunction, maintaining or using,
please contact Philips Consumer Care.
- Do not open, disassemble or repair the device
yourself. Modification of the device voids your
warranty.
- Report to Philips Consumer Care if any
unexpected operation or events occur.
- Dispose of accessories, detachable parts, and
the ME equipment according to the local
guidelines.
- Do not attempt to replace your blood pressure
monitor's battery. It is built-in and not
changeable.
- Avoid charging your blood pressure monitor in
extremely high or low temperatures (see
'Specifications').
- Do not clean the blood pressure monitor when
it is being charged. Always unplug the charger
first before cleaning the blood pressure
monitor.

Compliance with standards

- The device meets the relevant standards for this
type of Class IIa electrical medical equipment
for home use.
- This Philips device complies with all applicable
standards and regulations regarding exposure
to electromagnetic fields and complies with EN
60601-1-2.