Specifications - Philips DL8760 Manual

Upper arm blood pressure mobitor whith bluetooth smart
Hide thumbs Also See for DL8760:
Table of Contents

Advertisement

Problem
The display
shows Err
The display
shows E3
The display
shows E10
or E11
The display
shows E20
The display
shows E21
The display
shows EExx
Data
transmis­
sion or
pairing
failed.

Specifications

Product name
Power supply
Display
Measurement method Oscillometric method
Measurement range
Accuracy
Normal operating
condition
Storage and
transportation
conditions
Measurement
perimeter of the
upper arm
Net weight
External dimensions
Accessories
Mode of operation
Degree of protection
Protection against
ingress of water
Possible cause
Communication
error.
The cuff is not
properly
secured.
The device
detected
motion, talking
or the heart
rate is too weak
during the
measurement.
The device
does not detect
the heart rate
signal.
The
measurement
failed.
A system error
occurred.
Bluetooth is off. Turn on Bluetooth
The Philips
HealthSuite
health app is
off.
The blood
pressure
monitor and
mobile device
are more than
16 feet (5
meters) feet
away from
each other.
You selected
the wrong
profile on the
blood pressure
monitor.
Philips upper arm blood
pressure monitor with
Bluetooth® Smart
3.7V 1000mAH built-in
rechargeable li-polymer
battery, 6V 1A DC charger
Display with white
LED backlight
Visible area = 3 3/8" (L) x
15/16" (W) / 86.1 mm (L) x
24 mm (W)
Rated cuff pressure:
0mmHg - 300mmHg.
Measurement pressure:
40mmHg - 230mmHg
Heart rate: 40-199 beats
per minute
Pressure: 41°F to 104°F /
5°C - 40°C within ±
3mmHg. Heart rate: ±5%
of measurement result
on display
Temperature: 41°F to
104°F/ 5°C to 40°C.
Relative humidity:
≤85%RH. Atmospheric
pressure: 86kPa to
106kPa
Temperature: -4°F to
140°F / -20°C to 60°C.
Relative humidity: 10% to
93%. Atmospheric
pressure: 50kPa to
106kPa
About 8 ¾ - 16 ½ inch
(22- 42 cm)
Approx. 9.34 oz / 265g
Approx. 5 1/8" × 2 7/8" × 1
3/16"
(130.9mm × 73mm ×
29.4mm)
DC charger, user manual
Continuous operation
Type BF applied part
IP22, This means:
protected against access
to hazardous parts with a
finger and against
vertically falling water
drops when tilted up to
15 degrees.
Solution
Check if the app is
on and try data
transmission
again.
Refasten the cuff,
wait 3 minutes
and then measure
again.
Wait for 3
minutes and then
measure again.
Do not move
during
measurement.
Make sure the
device is in
contact with the
skin. Loosen the
clothing on the
arm and measure
again.
Wait for 3 minutes
and then measure
again.
Retake the
measurement. If
the problem
persists, call
1-844-531-6861
for assistance.
on your mobile
device.
Press the icon on
your mobile
device to activate
the health app.
Place your mobile
device closer to
the blood
pressure monitor.
Select the correct
user profile on the
blood pressure
monitor before
your
measurement.
Otherwise the
data cannot be
transmitted to
your app. Repeat
the measurement
with the correct
profile selected