Sourcing Loop Voltage

The Calibrator can supply loop power at 24 V dc to a
current transmitter that is disconnected from the system.
Use the following procedure:
1.
With power off, hold down
on. The "Loop Power" icon appears in the display.
2.
With the transmitter disconnected from normal loop
power, connect the Calibrator mA (+) and COM (-)
test leads in series with the instrument current loop as
shown in Figure 7.
3.
Measure loop current in the mA display.
O
4.
Press
off to deactivate the 24 V dc supply when
you are done sourcing loop voltage.
In addition to mA, the current can be displayed in two
alternative modes:
Percent Mode- The current is displayed as a
percentage based on a 4-20 mA scale.
Percent Error Mode- Transmitter current output error
is displayed. Error is calculated based on a
configurable zero and span pressure and a 4-20 mA
scale.
1.
Measure loop current in the mA measurement display.
O
2.
Press
OFF to deactivate the 24 V dc supply when
you are done sourcing loop voltage.
U
O
while pressing

Percent Error Setup

a
Press and hold
. After 3 seconds the set icon and 0%
appears on the lower display. Use
0% point for the Percent Error calculation, then press
D
(ENTER) key to confirm selection. Press
is displayed on the lower display. Use
the 100% point for the Percent Error calculation. Press
D
to confirm the selection and to exit.
(ENTER)
+
Figure 7. Sourcing Loop Voltage
Pressure Calibrator
Sourcing Loop Voltage
W
X
and
to adjust the
a
. 100%
W
X
and
to adjust
S I G N A L
T E S T
qo007f.eps
19