MANUAL - SOLVENT RECYCLERS - URS1600CE AND URS2000CE
PREPARATION AND SETUP:
•
Carefully inspect the shipping carton for any sign of transport damage.
•
Carefully remove the unit from the shipping carton.
•
Check the unit for damage. Report any transport damage immediately to the carrier and your
vendor. Initiate a freight claim with the carrier. The manufacturer is not responsible for freight dam-
age.
•
A Liner Bag and Retainer Ring are already installed inside the distillation tank.
•
Check the Accessory Kit for the parts listed below. If any parts are missing, contact your supplier. Ad-
ditional consumables and accessories are also listed.
ACCESSORY KIT CONTENTS
Manual
Lid Gasket
Liner Bag, 2 Spares
Filler Hose and Suction Pipe Assembly
Note: Liner bags are not included with the model, RI80.
LOCATION AND CONNECTION:
This unit is classif ed as Zone 1. Such equipment will not generate a spark. This equipment may be safely
placed in a Class 1, Division 1, Group D, Zone 1 locations where solvents may be present. See Hazard-
ous Location in the National Electrical Code for more information.
If this unit is placed in a Class 1, Division 2, Zone 2 or unclassif ed location consider the area within 10
feet from the unit as Class 1, Division 2, a location where solvent is sometimes present. All equipment in
this area, such as electrical and ventilation equipment must be certif ed for use in a Class 1, Division 2,
Zone 2 location. Such equipment cannot produce a spark. Equipment that can generate sparks, like gen-
eral plugs, welding equipment and power tools cannot be located in this area. Smoking is not permitted.
We recommend, as a best practice, that you keep this area open and free from all debris and equipment.
If you have a question contact your local f re prevention authorities for advice. The unit must be at least 1
foot from the back wall and 10 feet from 3 other walls.
The solvent recycler comes with an explosion proof plug. A qualifi ed electrician, in accodance
with the National Electrical Code, must install an explosion proof recepticle on the wall for con-
3
Revised 2017-10