Settings
Settings menu is where you can set the general
settings. These include Language, Beeper, Select
Units, Time Constant, Contrast, Set Time, Set
Date, Time Format, Date Format, Number Format,
Backlight, Auto Off. Use the arrow keys to select
an option or to change the settings for each
option. Press the
key to accept settings.
In addition to the general settings, the Settings
menu also includes RS232 Baud Rate, Deadband
and Headphone Input configuration settings.
RS232 Baud Rate
Deadband
Headphone Input
Use the arrow keys to select an option or to change the settings for each option. Press the
accept settings.
26
Sets the speed a computer will accept information.
The deadband determines the velocity or volume measurement threshold for
displaying 0. If the deadband is Enabled, readings below 18 fpm (0.0914 m/s)
3
or 18 cfm (30.58 m
/h, 8.49 l/s) will be displayed as 0. If the deadband is
Disabled, readings down to 0 will be displayed. The default is set to Enable.
The Headphone Input on the side of the meter can be either Enabled or
Disabled. When set to Enabled, the meter will default to Capture Hood mode
when the base switch is connected. When set to Disabled, the meter will not
respond to the base switch input.
Language
Beeper
Select Units
Time Constant
Contrast
Set Time
Set Date
Time Format
Date Format
Number Format
Backlight
Auto Off
RS232 Baud Rate
Deadband
Headphone Input
Airflow™ Instruments ProHood Capture Hoods PH730/PH731
SETTINGS
English
Disable
1 Second
09:14 AM
01/07/12
MM/DD/YY
XX,XXX.YY
Enable
Enable
Enable
key to
5
12 hr
Auto
9600