NEC MD212G3 Installation & Maintenance Manual page 22

Table of Contents

Advertisement

3. Select your desired confi guration by the (+) plus button or the (-) minus button.
4. After selecting your preferred CONFIGURATION mode, hold down the (+) plus button and the (-) minus button
simultaneously for over 2 seconds again. Then confi rm any change of confi guration in the "PRESENT STATE" OSD main
menu.
5. Click the (+) plus button or (-) minus button once to close the "PRESENT STATE" OSD main menu.
Performing the DICOM conformance test
This function is available only when the display has been calibrated to DICOM GSDF.
Performing a DICOM conformance test ensures DICOM conformance of the display by measuring and checking the display
properties.
The DICOM conformance test is recommended to be performed after 60 minutes of continuous active backlight time of the LCD
panel. If the test is performed within less than 60 minutes of active backlight time, the reported DICOM GDSF error value may
be larger because the luminance was not yet stable enough.
HINT:
If the power management setting of the computer is set to less than 60 minutes, either change the computer's power
management setting only for this test or try "OSD operation without signal input from computer" on page 21.
DICOM conformance test procedure:
1. Checking the gamma setting
The (+) / (-) buttons are located in the back side of the display. Press either button twice, and the "PRESENT STATE" OSD
main menu appears.
Make sure that GAMMA is set to "DICOM GSDF".
Present operating state
P R E S E N T
S T A T E
L U M I N A N C E
:
5 0 1
A M B
L I G H T
:
3 0 0 lx
I N P U T
:
D i s p l a y P o r t
C A L I B R A T I O N
S E T T I N G
C O N F I G
:
1
L U M I N A N C E
:
5 0 0
G A M M A
:
D I C O M
C O L O R
:
6 5 0 0 K
English-19
G S D F
Gamma setting

Advertisement

Table of Contents
loading

This manual is also suitable for:

Md211g5

Table of Contents