It is strongly recommended that the following checks are carried out on a weekly basis by
trained personnel, to ensure trouble free operation of the trolley.
Check the solid plastic table top for cracks or other obvious signs of damage or
contamination. Report any damage found.
Check mattress for any sign of damage or contamination. Report damage or clean where
necessary with mild soap and water.
Check castors for free movement and the braking mechanism for functionality, check and
remove any debris that may have been caught between the forks and around the axles,
This is a common occurrence and can lead to a shortened life expectancy of the castor if
left unattended.
Check the backrest function.
Check the cot sides for functionality and damage, especially the cot side brackets and
plungers, as these take the brunt of any impacts from the side. Ensure that the plungers
work freely and locate into the holes.
Ensure that the IV Pole fitted is correct and the serial no. corresponds with the trolley serial
no. located on the inside of the backrest board (see page 5 for location). Make sure there
are no sharp burrs from misuse.
Please register your product with us, via the Warranty Registration Form (see page 13), for ease
of contact in case of urgent upgrades or recalls.
If the unit is damaged please do not attempt to repair as this could invalidate your
warranty and could compromise its MR integrity and therefore safety, please call or
email Wardray Premise Limited: