Barco Eonis MDRC-2224 User Manual page 32

24-inch clinical display
Table of Contents

Advertisement

7. Important information
Immunity test
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption and reflection from structures, objects and people.
Recommended separation distance
The Eonis is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer of the user of the Eonis can help prevent electromagnetic interference by main-
taining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the Eonis as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter
W
0.01
0.1
1
10
100
3. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Eonis is used exceeds the applicable RF compliance level above, the Eonis should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Eonis.
4. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
5. For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter. Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
30
IEC 60601
Test levels
Separation distance according to frequency of transmitter
5
150kHz to 80MHz
d=1.2√P
0.12
0.38
1.2
3.8
12
Compliance level
80MHz to 800MHz
d=1.2√P
0.12
0.38
1.2
3.8
12
K5903081 EONIS 08/03/2016
Electromagnetic
environment –
guidance
separation distance in
meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
3
survey,
should be less
than the compliance
level in each frequency
4
range.
Interference may occur in
the vicinity of equipment
marked with symbol:
800MHz to 2.5GHz
d=2.3√P
0.23
0.73
2.3
7.3
23

Advertisement

Table of Contents
loading

This manual is also suitable for:

K9303005a

Table of Contents