Problem
The picture is not in focus.
The DP2-TWAIN or cellSens /
OLYMPUS Stream window is not
displayed correctly or the menu
characters are not displayed correctly.
Another 1394 device cannot be used
simultaneously.
Cause
The microscope is not focused
properly.
The parfocality is not adjusted
properly.
The aperture iris diaphragm of the
condenser is open too wide.
The field iris diaphragm is not set
properly.
Lens components of the microscope
are contaminated or the cover glass
on the front of the camera is stained.
The microscope and/or camera
are subjected to vibration during
recording.
The resolution setting of the screen is
incorrect.
The large font has been selected for
the font size of the screen.
The DP26-dedicated driver is installed. When the DP26 is connected to a PC,
Remedy
Adjust the focus correctly with the fine
adjustment knob.
Adjust parfocality with the camera
adapter.
Close the aperture iris diaphragm a
little.
Adjust the field iris diaphragm until the
image circumscribes the field of view.
Clean the objective, photography lens,
condenser and/or window lens of the
microscope, or clean the cover glass
on the bottom of the camera head.
Record images in an environment in
which the microscope and camera
are not vibrated. It is effective to use an
anti-vibration bench.
Set the resolution setting at 1280 x
1024 or more in the properties of the
screen.
Select a small font in the properties of
the screen.
other 1394 devices cannot be used on
the same PC.
DP26
Page
6
10,21
28