What To Do, If; Que Faire, Quand - Xylem ebro EBI 20 Series Manual

Table of Contents

Advertisement

English
Close the battery compartment
(cover engages).
NOTE:
When changing the battery or if the bat-
tery fails, stored data are not lost. How-
ever, in a prolonged power failure the
time will no longer coincide with the set
time. The data logger must then be read
and reprogrammed and the date and
time resynchronized.
After inserting the new battery recording
and, if necessary, alarm signaling, will
continue.

What to do, if...

... the following messages appear on
the display panel:
Possible cause
Measurement
range exceeded
Sensor defective
Below
measurement
range
Sensor defective
Ba80019def02_EBI20
Remove fault
Observe
measurement
range
Call service
Observe
measurement
range
Call service
Français
Fermez la compartiment à batterie
(le couvercle s'emboîte).
REMARQUE:
Les données enregistrées ne s'effacent
pas en cas de défaillance ou de chan-
gement de la batterie. En cas de période
prolongée sans alimentation en courant,
l'heure indiquée par l'horloge interne
n'est plus correcte. Les données de l'ap-
pareil d'enregistrement doivent alors
être saisies et il doit être reprogrammé,
ce qui synchronisera à nouveau la date
et l'heure.
Après le mise en place de la nouvelle
batterie, l'enregistrement et le cas
échéant l'affichage de l'alarme repren-
dront.

Que faire, quand...

... le transfert de données par radio
vers l'enregistreur ne fonctionne
pas:
Cause posible
Plage de mesure
dépassée
Sensor defective
Plage de mesure
non atteinte
Senseur
défectueux
Remedy
Respectez la
plage de mesure
Appelez le S AV
Respectez la
plage e mesure
Appelez le S AV
27

Advertisement

Table of Contents
loading

Table of Contents