Solar Cell I-V Test System (Manual)
3.
Safety
Do NOT connect external voltage sources to either SMU channel.
•
The absolute maximum input voltage for the Vsense channels is ±12 V. Do NOT apply
•
input while not powered.
3.1
Use of Equipment
The Ossila Solar Cell I-V Test System (Manual) is designed to be used as instructed. It is intended
for use under the following conditions:
Indoors in a laboratory environment (Pollution Degree 2).
•
Altitudes up to 2000m.
•
Temperatures of 5°C to 40°C; maximum relative humidity of 80% up to 31°C.
•
The unit is supplied with a 24 VDC power adapter with a power cord for the country of purchase,
in accordance with European Commission regulations and British Standards. Use of any other
electrical power cables, adaptors, or transformers is not recommended
3.2
Hazard Icons
The following symbols can be found at points throughout the rest of the manual. Note and read
each warning before attempting any associated operations associated with it:
Table 3.1. Hazard warning labels used in this manual.
Symbol
Electrical shock
3.3
General Hazards
Before installing or operating the Ossila Solar Cell I-V Test System (Manual) there are several health
and safety precautions which must be followed and executed to ensure safe installation and
operation.
Ossila.com
Associated Hazard
6
User Manual
Ossila Limited © 2022