15. Troubleshooting Guide ........................11 16. Warranty ............................12 Thank you for purchasing Apera Instruments PH60 Premium pH Tester. Please read this manual before use in order to properly use and maintain the product. For video tutorials, please go to support.aperainst.com ATTENTION Ը...
3. Preparation before Use 3.1 Pull out the battery insulation slip, and take off the probe cap. 3.2 Rinse off the probe in pure water (preferably distilled or deionized water. RO water or tap water is the alternative) , then shake off excess water. 3.3 Perform at least a 2-point calibration.
Page 5
4.1.5 Wait for the reading to stabilize (when stays on the screen), then short press to finish the first point calibration. The tester returns to measurement mode and icon will appear at the bottom left, indicating a successful 1-point calibration (the middle point). 4.1.6 To calibrate second point, use pH 4.00 buffer and repeat Step 4.1.3 to 4.1.5 (Do NOT turn off the tester after you finish pH 7 calibration).
5. pH Measurement 5.1 How to take pH measurements Short press to turn on the tester. Rinse the probe in pure water, shake off excess water. Insert the probe in your sample solution, make a quick stir and hold still. Record the reading when it is stabilized ( comes up and stays on screen).
8. Parameter Settings 8.1 Setup Menu Factory Symbol Parameter Setting Contents Code Default Select pH buffer solution USA – NIST Low value measurement alarm setting 0 to 14.00pH High value measurement alarm setting 0 to 14.00pH 14.00 Auto. Hold Off – On Select backlight Off - 1 - On Select temperature unit...
8.3.6 Factory default setting (P7): Select “Yes” to set the meter to its default status (erase all calibration record and return all parameter settings to the default value). This function can be used when the meter does not work well in calibration or measurement. Calibrate the meter again after setting the meter to factory default.
1-2 years depending on many factors such as frequency of use, nature of test samples, and how well it is maintained, etc. Apera Instruments recommends replacing the pH probe every 1 to 2 years to guarantee the optimal performance.
13. Notes Avoid testing in very high (>113°F) or very low (<41°F) temperature solutions as it will cause greater measurement error and potential damage to the probe. Test your samples and perform calibration close to room temperature as much as possible. b) NEVER store the probe in pure water like tap, RO, distilled, or deionized water as they could damage the pH probe.
15. Troubleshooting Guide Trouble Potential Reasons How to Fix When powered on, calibrate pH 7 first, then pH 4. After pH 4 is Incorrect calibration order calibrated, if you want to calibrate pH7 again, you need to power off and restart the tester. Bad standard solutions Replace with a new reliable buffer solution.
We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at option of APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS, LLC for a period of TWO YEARS (SIX MONTHS for the probe) from the delivery.