Function checks
Checking the exposure control
•
Check the exposure control with the following setting:
•
Using a clock, check the acoustic signal for the exposure time.
Proceed as follows:
➪ Release an exposure; the radiation indicator (9/Fig. 5) lights up.
➪ The exposure must switch off after max. 6s.
➪ The radiation indicator goes out, the acoustic signal stops.
Custom programming
•
The maximum possible kV and mAs values can be programmed with service program
Pr.6 (refer to the service instructions "Service programs").
•
The default exposure parameters can be programed with service program Pr.5 (refer to
service instructions "Service programs")
Checking coincidence of light and radiation fields
•
Load a 24 cm x 30 cm or 10"x 12" cassette with film and place it on a table or a similar
surface.
•
Set the vertical SID to 100 cm or 40" with the measuring tape reaching to the upper edge
of the cassette.
•
With the control buttons set a format of 18 cm x 24 cm or 8"x 10".
•
Switch on the light localizer and align the cassette.
•
Place radiolucent markers (e.g. washers, coins) on the cassette as shown in the sketch.
Use one washer as side marking.
•
Release an exposure (60 kV, 10mAs) and develop the film.
•
On the developed film, write the following data with a waterproof felt pen:
- SID setting
- film size
- radiation field size