Affirm Breast Biopsy Guidance System User Guide
The order within a Multi-Pass target starts at the center target point. After
the center target point, the order moves to the 12 o'clock position and
continues counter-clockwise through the offset points. See the following
figures.
Figure 37: Biopsy Order
of a Three Offset Target
Verify the position of the Biopsy Device (refer to
on page 38). If necessary, observe the targets on post-Tomosynthesis scout images
(refer to
Project Targets on Post-Fire Scout Image
Put in a biopsy site marker, if desired.
Move the biopsy device away from the breast.
Acquire images as necessary.
Release compression.
Figure 38: Biopsy Order of a
Four Offset Target
Verify the Position of the Biopsy Device
on page 39).
Chapter 5: Biopsy
Figure 39: Biopsy Order of
a Five Offset Target
Page 43