44 | CR 30-X/CR 30-Xm Plates and Cassettes | Cleaning
• Take an appropriate wipe or moisten the cloth with cleaner and wipe
• Leave the image plate surface to dry for about 10 minutes to allow the
• Use only a dry cloth or blow away possible dust with compressed air
• Once the image plate surface is dry (after about 10 minutes), check
7. Put the tray with the image plate back into the cassette.
1.
2.
8. When the black tray with the image plate is completely inside the cassette,
close the shutter.
To close the shutter easily, insert a key or a pen into the cassette lock while
closing the shutter. Once closed, remove the key or the pen.
2387E EN 20180416 1238
the image plate surface softly and evenly.
solvents to evaporate.
(canned air), to clean the tray thoroughly. (DO NOT USE A SCREEN
CLEANER TO CLEAN THE TRAY.)
once again for particles of material and other impurities before placing
the tray and image plate set into the cassette.
WARNING:
Verify that the white phosphor side is oriented to the black
tube side of the cassette and that the shutter does not scratch
the image plate.
White phosphor side
Black tube side of the cassette
CAUTION:
Make sure to slide in the image plate carefully. Do not let the
plate fall down vertically into the cassette! This causes
damages of the phosphor.