CR 30-X/CR 30-Xm Plates and Cassettes | CR 30-X/CR 30-Xm Plates and Cassettes Description | 25
1
6
Label
1.
Clip
2.
Cassette format
3.
Yellow dot
4.
Lock to open and close the cassette
5.
Shutter-opening mechanism
6.
Figure 4: Layout
CR FLFS cassettes are specifically designed for the Full Leg Full Spine
application and are easily distinguishable from CR General cassettes by its
yellow labeling and yellow dots.
Limitations
The use of CR General cassettes for Full Leg or Full Spine imaging will result in
a white line in the stitching zone, in which there is no image information.
CR FLFS cassettes can be used for other than Full Leg or Full Spine imaging,
however with a limitation in the border zone. Due to the presence of the
reduced backscatter protection foil at the 35 cm borders of the cassette, there
is a risk for reduced image quality in this border area (of maximum 1 cm) if
this zone was exposed and in case there was an influence of backscattered
radiation.
Risk of reduced image quality in border areas:
Note: In the CR 30-X you cannot use other than the above described
cassettes.
2
3
4
5
2387E EN 20180416 1238