DR Generator Sync Box, DR Retrofit Solution | Remarks for HF-emission and immunity | 57

Remarks for HF-emission and immunity

It is hereby certified that the DR Generator Sync Box has interference
suppression according to the EN 55011 Class A as well as the FCC Rules CR47
Part 15 Class A.
This device was tested for a normal hospital environment as described above.
The user of the device should ensure that it is used in such an environment.
This equipment has been tested and found to comply with the limits for a
Class A digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference when
the equipment is operated in a commercial environment. This equipment
generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the user
will be required to correct the interference at his own expense.
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
RF Emission Meas-
urements
High frequency RF
emissions in ac-
cordance with
CISPR 11
High frequency RF
emissions in ac-
cordance with
CISPR 11
WARNING:
This device is intended for use by healthcare professionals only.
This device may cause radio interference or may disrupt the
operation of nearby equipment. It may be necessary to take
mitigation measures, such as re-orienting or relocating the
device or shielding the location.
WARNING:
The HF-emission and immunity can be influenced by connected
data cables depending on length and the manner of installation.
Agree-
ment
Group 1
Class B
Electromagnetic Environment Guide-
lines
The device uses high frequency energy ex-
clusively for its internal functions. For this
reason, its high frequency RF emission is
very low and it is improbable that neighbor-
ing electronic equipment will be disrupted.
The DR Generator Sync Box is intended for
use in all buildings, including living areas
and areas directly connected to a public
supply network that also supplies buildings
that are used for domestic purposes.
0319D EN 20191031 1011