26 | DR Generator Sync Box, DR Retrofit Solution | Introduction

Connectivity

The DR Generator Sync Box is connected to the NX workstation and the X-ray
generator and replaces the interface to the original exposure button.
The DR Generator Sync exposure button is connected to the DR Generator
Sync Box device.
On supported X-ray systems, the NX workstation is connected to the X-ray
system to exchange X-ray exposure parameters.
The NX workstation requires a 100 Mbit ethernet network to exchange
information with a number of other devices.
The NX workstation communicates with other devices in the hospital network
using one of the following protocols:
• DICOM
• IHE
The NX workstation can be connected to a RIS system (input scheduling), a
PACS system (output image/data management) and to a hardcopy device
(output image).
Related Links
Configuration

Connectivity requirements

0319D EN 20191031 1011
Note: The connections between the components of the DR
Generator Sync Box and towards the NX workstation and the X-ray
system should not be disconnected or modified.
on page 12
Warning: Any kind of equipment connected to any interface
must be certified according to the respective IEC standards
(e.g. IEC 60950 / IEC 62368 for data processing equipment
or IEC 60601-1 for medical equipment). Furthermore all
configurations shall comply with the requirements for
medical systems according to IEC 60601-1. Everybody who
connects additional equipment to the signal input part or
signal output part configures a medical system, and is
therefore responsible that the system complies with the
requirements of the valid version of the system standard IEC
60601-1. If in doubt, consult your local service organization.