Step 2: select the exposure

In the operator room:
1. At the NX workstation, select the thumbnail for the exposure in the Image
Overview pane of the Examination window.
The selected DR Detector is activated.
The DR Detector Switch shows which DR Detector is active and shows its
status.
• Red (flashing): starting up
• Green (constant): ready for exposure
2. At the X-ray generator console or on the DR Software Console, select the
exposure settings suitable for the exposure.
On systems with integration of X-ray exposure parameters, the default X-
Ray exposure parameters for the selected exposure are sent to the
modality and displayed on the DR Software Console.
DR Generator Sync Box, DR Retrofit Solution | Getting started | 41