Disconnecting the DR Generator Sync Box from the mains
power
To disconnect the DR Generator Sync Box from the mains power, switch off
the power switch or remove the power plug from the socket.
DR Generator Sync Box, DR Retrofit Solution | Introduction | 37
Warning: The DR Detector or CR cassette are not intended
to be used as a primary barrier to X-rays. The user is
responsible for ensuring the safety of the operator,
bystanders, and the subjects being radiographed.
Warning: Avoid unnecessary dose by checking before
exposure if the DR Detector Switch displays the name of the
DR Detector that is being used and if the status of the DR
Detector is ready for exposure.
Warning: Avoid unnecessary dose by checking the
workstation selection on the X-ray generator console before
exposing. In a configuration with a DR Detector configured
on a virtual port, the DR Detector will not be triggered if a
free exposure is selected on the Generator console and yet
the exposure will be allowed.
Warning: To prevent unintended exposure, the position of
the exposure footswitch should be such that it cannot be
accidentally stepped on.
Caution: Excessive ambient temperature may impact
performance of DR Detectors and cause permanent damage to
the equipment. If ambient temperature and humidity is outside
the range specified in the technical data, do not operate the
system or use air conditioning. Warranty will be void if it is
obvious that operating conditions are not met.