Installation - ELGA MEDICA Pro EDI Quick Start Manual

Hide thumbs Also See for MEDICA Pro EDI:

Advertisement

ELGA
MEDICA Pro EDI
Maintenance
Electrical supply
Site
Water supply
3.2 External services and connections
Connection
Feed water inlet*
Pressurized drain
EDI drain
Application loop feed
Application loop return
Application feed
Electrical supply
* Refer to Operator manual for feed water specification. Suitability of water supply must be confirmed before commissioning is commenced.
Page 2
Access is required to the front of the unit for regular maintenance procedures.
Occasional access is required to the top of the unit to add chemical during sanitization.
For safety reasons it is recommended that this unit be connected to a suitable power
outlet which is easily accessible and local to the unit.
The unit can be floor or bench mounted. Ensure the bench is designed to take a load in
excess of 150kg and that the weight of other equipment is also taken into consideration.
Local potable supply with a means of isolation.
Termination
10mm OD Push fit
10mm OD Push fit
10mm OD Push fit
12mm OD Push fit
12mm OD Push fit
12mm OD Push fit
IEC 60320
MEDICA Pro EDI MANU40683 Ver.2 – 10/15
3.

INSTALLATION

3.1 Unpacking MEDICA Pro EDI -US
The following items are supplied:
1.
MEDICA Pro. EDI - US
2.
Electrical supply lead.
3.
Quick reference guide fitted on inside of door.
4.
MEDICA Pro Starter Kit 6 or 7 (to be ordered
separatly).
5.
Installation kit (Included in starter kit).
Specification
Flow
Pressure
Temperature
Maximum flow
Maximum
distance to drain
Maximum back
pressure
Maximum pipe
length to
application
Maximum pipe
length from
application
Maximum pipe
length to
application
Electrical rating
Electrical
protection
MEDICA Pro EDI - US Quick Start Manual
UNIT OF MEASURE
METRIC
US/IMPERIAL
>9l/min
>2.4USG/min
2 to 6 bar
30 to 88psi
4 to 40C
39.2 to 104F
10l/min
2.7USG/min
3m
10ft
0.2bar
3psi
5m
16ft
5m
16ft
3m
10ft
208-230 V ac 50/60 Hz
100-115 V ac 50/60 Hz
10 A