24
• When removing the output connector, release the lock
and then grasp the connector and pull to remove. Never
try to forcibly remove the connector without first
releasing the lock. Also, never pull the cable, as this
may damage the terminator.
• When connecting this instrument to an input terminal
that is not a BNC terminal, make sure that the polarity is
correct.
• Do not demagnetize while this instrument is clamping an
insulating target conductor to be measured.
Demagnetizing causes current to flow into the insulating
target conductor to be measured, which may damage
parts in the circuit to be measured.
Demagnetizing
• Check that the insulating target conductor to be
measured is not clamped when supplying power to this
instrument
waveforms may generate when supplying electric
power.
Chapter 2 Operation
Insulating target conductor
for
the
Cautions
same
reason.