Advertisement

Quick Links

mySense
Instructions for use
www.cyclotest.de

Advertisement

Table of Contents
loading

Summary of Contents for uebe cyclotest mySense

  • Page 1 mySense Instructions for use www.cyclotest.de...
  • Page 2: Table Of Contents

    Table of contents Intended use Intended use The cyclotest mySense is a medical digital thermometer (direct ® mode) for the self-measurement of human basal body temperature. Safety instructions The device transmits the measured values to the cyclotest mySense ® Operating the device app via a Bluetooth 4.0 interface.
  • Page 3: Safety Instructions

    Safety instructions Safety instructions • The device must be used solely for the purpose described in these • If the device has been stored at temperatures below 10°C or above operating instructions. The manufacturer shall not be held liable for 40°C, allow it to stand at room temperature for at least 2 hours be- any damages resulting from improper use.
  • Page 4: Operating The Device

    Operating the device Operating the device Device description Display Temperature unit Memory mode Gold-plated measuring tip On/off button Bluetooth data connection Synchronising Data transfer and memory Battery compartment Battery display Time recall button Date Illuminated display Initial device setup You must install the cyclotest mySense app on your smartphone be- ®...
  • Page 5: Temperature Measuring

    Operating the device Operating the device Open the cyclotest mySense app and complete the user registration. To perform the test measurement, please follow the instructions in ® Follow the instructions in the app. the app and refer to the instructions for measuring wake-up temper- ature starting on page 10.
  • Page 6 Operating the device Operating the device • If more than 25 saved measurements have • Place the probe tip underneath your tongue on either the left-hand not been transferred to the mySense App, the or right-hand side. The temperature sensor must make good con- message “APP/FULL“...
  • Page 7: Manually Transferring Data To The Mysense App

    Operating the device Operating the device • If the connection is successful, the “APP“ indi- If the device cannot connect to the mySense app, it switches itself off cator appears and the display lights up green after 60 seconds. If this occurs, check whether Bluetooth is enabled on (Figure 11).
  • Page 8: Battery Display/Changing The Battery

    Operating the device Operating the device • By repeatedly pressing the memory recall • Place the cover back on the battery compartment and lock it in , you can call up the previous measure- place. ments (no. 2, no. 3 ...). After changing the battery, the thermometer must be connected to •...
  • Page 9: Technical Information Error Messages

    Technical information Technical information Error messages No display after switching on. Check whether the bat- tery is positioned correctly and the polarity (+/-) is cor- The measured temperature is above 43.99°C and is rect. If the error persists, change the battery. therefore outside the measuring range.
  • Page 10: Customer Service

    We recommend a technical inspection of the device every 5 years. This racy: ± 0.10°C in other temperature ranges inspection can be carried out either by UEBE Medical GmbH or by au- thorised maintenance services. Memory: 30 measurement values with date and...
  • Page 11: Cleaning And Care

    In the case of a warranty claim, the device along with the proof of purchase and a description of the complaint must be sent to: UEBE Medical GmbH Service-Center Zum Schlag 18 97877 Wertheim, Germany The cost of returning the device to our factory shall be borne by the sender.
  • Page 12: Cyclotest ® Service Hotline

    cyclotest service hotline ® You have undoubtedly read our instructions on the use of cyclotest ® mySense carefully and have familiarised yourself with its character- istics. If you have any additional questions, you can call our cyclotest ser- ® vice hotline, which is available Monday to Thursday from 8:00am to 5:00pm and Fridays until 4:30pm.
  • Page 13 Phone: + 49 (0) 93 42 / 92 40 40 Fax: + 49 (0) 93 42 / 92 40 80 Email: info@uebe.com Website: www.uebe.com Subject to technical modifications. Reprinting, including of extracts, is prohibited. © Copyright 2019 UEBE Medical GmbH 0123...

Table of Contents