Summary of Contents for Kane 3500

  • Page 1 KANE3500 Pressure Meter Stock No: 18854 -2 June 2014 © Kane International Ltd...
  • Page 2: General Operation

    General Operation When the instrument is first switched on using , the top line of the display shows in sequence the date, the time, the model version number, the battery status and the pressure units. The bottom line of the display shows a countdown number that starts at 10 and reduces to zero.
  • Page 3 UP / DOWN Press either to change the bottom line of the display. ENTER Press to cancel OFF . The buttons also change settings such as date, time, and other menu items when MENU mode has been selected. KANE3500 Manual Page 3...
  • Page 4: Before Using The Meter For The First Time

    1. BEFORE USING THE METER FOR THE FIRST TIME: Remove the cover and fit a new battery in the battery compartment. Take great care to ensure that the battery is fitted with the correct battery polarity. Then replace the battery cover. Always check that the meter is working correctly after replacing the battery. Set the meter’s correct time, date and units.
  • Page 5 The final, logical returns you to the main menu display. To exit the MENU function the final logical is pressed. No changes are made if the MENU function is aborted by pressing PRESSURE SMOOTH OFF = Normal ON = Readings filtered to damp out fluctuations RESOLVE HIGH = maximum number of decimal places LOW = one less decimal place displayed...
  • Page 6 STORE VIEW: The test number appears on the top line. to change test number. The bottom line shows the measured value. to exit this mode. LOG : Saves a set of results when is pressed. Storage capacity is 255 sets of results AUTO STO : Automatically stores readings at a preset interval.
  • Page 7 Max Pressure is displayed as: Min Pressure is displayed as: MEASURING Make sure you do not exceed the meter’s operating specifications.  Do not exceed the meter’s internal temperature operating range  Do not put the meter on a hot surface When taking critical draft measurements always re-zero the meter in the position you are taking the readings and hold the meter still during the test.
  • Page 8 Always use the top right hand port (Port 1) for taking single channel measurements (pressure or vacuum). Only use the left hand port for differential measurements. If the pressure being measured exceeds the meter’s design range the display will show OR for “over-range”...
  • Page 9: Meter Annual Recalibration And Service

    5. METER ANNUAL RECALIBRATION AND SERVICE The meter should be re-calibrated and serviced annually by a Kane approved service centre. Local regulations may require more frequent re-calibration. Calibration is performed in firmware and there are no user accessible adjustments or user serviceable parts.
  • Page 10 Model: KANE3500-5 Nominal Pressure Ranges mBar 400.0 4000 32000 40.00 mmHg 5.000 300.0 400.0 150.0 Model: KANE3500-15 Nominal Pressure Ranges mBar 1040.0 100.00 15.00 mmHg 780.0 1040.0 400.0 Model: KANE3500-30 Nominal Pressure Ranges mBar 2070.0 207.00 30.00 mmHg 1560.0 2070.0 800.0 KANE3500 Manual Page 10...
  • Page 11 Pressure Measurement Model: KANE3500-1 Max. overrange without damage to sensor is 5 times nominal range Range Resolution Accuracy + 20 Pa 0.1 Pa +0.5 Pa + 100 Pa 0.1 Pa + 3 Pa + 2000 Pa 1 Pa + 3% of reading + 80 hPa 0.01 hPa + 3% of reading...
  • Page 12 Model: KANE3500-15 Max. overrange without damage to sensor is 5 times nominal range Range Resolution Accuracy + 309.99 hPA 0.01 hPa + 10 hPa + 1040.0 hPA 0.1 hPa + 3% of reading Model: KANE3500-30 Max. overrange without damage to sensor is 5 times nominal range Range Resolution Accuracy...
  • Page 13: Electromagnetic Compatibility

    If not acceptable, adjust the meter’s position to minimize interference or switch off, if possible, the offending equipment during your test. At the time of writing this manual (June 2014) Kane International Ltd are not aware of any field based situation where such interference has occurred and this advice is only given to satisfy the requirements of the Directive.
  • Page 14: Appendix 1 - Main Parameters

    APPENDIX 1 – Main Parameters : User selectable from : UNITS : inH2O mBar mmH2O NOTE! Not all units available on all models. DATE : Date shown as day, month and year. The order can be changed using the menu function.
  • Page 15 Once your KANE3500 has been securely packed then your package is ready for shipment back to Kane. If you do not have an account with a courier company you can take your package to your local Post Office. It is advisable to send the package by Special Delivery so that it is insured and traceable while in transit.
  • Page 16 Service Returns (Simply cut out and attach to your package) Southern Service Centre Kane International Ltd Kane House, Swallowfield Welwyn Garden City Hertfordshire AL7 1JG Southern Service Centre Kane International Ltd Kane House, Swallowfield Welwyn Garden City Hertfordshire AL7 1JG...
  • Page 17: Product Registration

    Product Registration Please complete, detach and return to: Kane International Ltd, Kane House, Swallowfield, Welwyn Garden City, Hertfordshire, AL7 1JG Your Details Name: Job Title: Company Name: Company Address 1: Address 2: Town/City: County: Postcode: Country: Phone Number: Fax Number:...
  • Page 18 Why did you buy a Kane Product? Dealer Recommendation Other Recommendation   Value for Money Our Fixed Price Servicing Programme   Kane Previous Owner   Not your Decision Other:   What brand was your previous analyser?
  • Page 19 KANE3500 Manual Page 19...
  • Page 20 Thank you for buying this pressure meter. Before use, please register on our website www.kane.co.uk or complete, detach and return the Product Registration page. KANE3500 Manual Page 20...

Table of Contents