Specifications: Technical Data - Withings BPM Core Instruction Manual

Table of Contents

Advertisement

Available languages

Available languages

Specifications: technical data

EN
Product description
Model
Blood pressure measurement
method
Cuff inflation
Pressure sensor
Measurement range (pressure)
Measurement range (pulse)
Pressure sensor accuracy
Clinical accuracy (blood pressure)
Automatic electronic blood pressure monitor with stethoscope, electrocardiogram and heart rate sensor
WPM04
Cuff oscillometric method
Automatic inflation with air pump at 5 mmHg/s
Gauge sensor
0 to 285 mmHg, DIA 40 to 130 mmHg, SYS 60 to 230 mmHg
40 to 180 beats/min
Within +- 3 mmHg or 2% of reading
BPM Core performance in blood pressure measurement was validated in a clinical trial. The results are
within the margin of acceptance defined by the internationally recognized evaluation standard of blood
pressure monitors ANSI/AAMI/ISO 81060-2:2013, EN ISO 81060-2:2014, developed by the European
Society of Hypertension, British Hypertension Society and Association for the Advancement of Medical
Instrumentation/American Heart Association. For more information, go to: withings.com/support

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents