2645A Four-Wire Resistance Measurement Specifications

Tables 1-34 to 1-36 provide 2645A specifications for the four-wire resistance
measurement function. The four-wire measurements use 2 input channels a decade apart,
e.g., channels 4 and 14.
Table 1-34. 2645A Four-Wire Resistance Temperature Coefficient
Specification
Temperature
Coefficient
Table 1-35. 2645A Four-Wire Resistance Range and Resolution Specifications
Resolution
Range
Slow
300Ω
10 mΩ
30 mΩ
3 kΩ
100 mΩ
300 mΩ
30 kΩ
1Ω
3Ω
300 kΩ
10Ω
30Ω
3 MΩ
100Ω
300Ω
Table 1-36. 2645A Four-Wire Resistance Accuracy Specifications
Range
90 Day
Slow
300Ω
.02%+60 mΩ
.02%+.1Ω
3 kΩ
.02%+.6Ω
.02%+2Ω
30 kΩ
.02%+6Ω
.2%+200Ω
300 kΩ
.5%+80Ω
1%+2 kΩ
3 MΩ
1.3%+1 kΩ
2%+120 kΩ

2645A Two-Wire Resistance Measurement Specifications

The 2645A specifications for the two-wire resistance measurement function is based on
the four-wire resistance measurement specification (above) except you add a 700 to 1000
ohm positive offset. This value varies for each channel and temperature gradient
(nominal +1%/ºC).
Add 1/10th the 90 day specification per °C above 28°C or below 18°C.
Current
Applied
Fast
1 mA
300 mV
100 µA
300 mV
10 µA
300 mV
10 µA
3.0V
1 µA
3.0V
Accuracy, 3σ + (% input + Ω)
18°C to 28°C
1 Year
Fast
Slow
.02%+.1Ω
.02%+1Ω
.02%+10Ω
.5%+150Ω
1.3%+2 kΩ
Introduction and Specification
Characteristic
Full Scale
Maximum Voltage
Voltage
Applied by Instrument
3.5V
3.5V
3.5V
3.5V
3.5V
-10°C to 60°C
1 Year
Fast
Slow
.02%+.2Ω
.084%+.25Ω
.02%+3Ω
.084%+2.5Ω
.2%+300Ω
.084%+25Ω
1%+3 kΩ
2.1%+336Ω
2%+200 kΩ
5.46%+4.2 kΩ
Specifications
1-32.
Fast
.084%+.42Ω
.084%+8.4Ω
.84%+840Ω
4.2%+8.4 kΩ
8.4%+200 kΩ
1-33.
1
1-23