2640A/2645A General Specifications - Fluke 2640A Service Manual

Netdaq data acquisition tools
Hide thumbs Also See for 2640A:
Table of Contents

Advertisement

2640A/2645A General Specifications

Table 1-5 provides the general specifications for the 2640A and 2645A instruments.
Specification
Channel Capacity
20
I/O Lines Total
12
Size
9.3 cm (3.67 in) high, 21.6 cm (8.5 in) wide, 36.2 cm (14.28 in) deep
Weight
Net, 4 kg (8.8 lb.) Shipping, 6.0 kg (13.2 lb.)
Power
107 to 264V ac (no switching required), 45 to 65 Hz, 15 VA maximum
9V dc to 16V dc, 6W maximum. Specifications are for 50 or 60 Hz operation.
If both sources are applied simultaneously, ac voltage is used if it exceeds
approximately 8 times the dc voltage. Automatic switchover occurs between ac
and dc without interruption.
Standards
Both instruments comply with:
Serial Interface
Connector: 9 pin male (DB-9P)
(RS-232C)
Signals: TX, RX, DTR, RTS, GND
Modem Control: full duplex
Baud rates: 4800, 9600, 19200, 38400
Data format: 8 data bits, no parity bit, one stop bit
Flow control: XON/XOFF
Echo: Off
Common Mode
2640A 150V (300V on channels 1 and 11)
Voltage
2645A 50V dc or 30V ac rms.
Measurement Speed
2640A
(Scanning Rates)
2645A
Accuracy of Medium
Equal to (Fast Accuracy Rate + Slow Accuracy Rate)/2
Scanning Rate
Additional error if
If the instrument is fully warmed-up at the time drift correction was disabled, i.e.,
turned on at least 1 hour earlier: 1/10 of the 90-day specification per °C change
"Automatic drift
correction" is turned
in ambient temperature from the temperature when drift correction was disabled.
off.
If the instrument was not fully warmed up at the time of drift correction was
disabled: add an error equal to the 90-day specification for instrument warmup
+1/10 of the 90-day specification per °C change in ambient temperature from the
temperature when drift correction was disabled.
Table 1-5. 2640A/2645A General Specifications
IEC 1010-1
UL 1244
CSA Bulletin 556B
ANSI/ISA-S8201-1988
CSA C22.2 No. 101.1-92
Vfg. 243/1991 (when shielded cables are used)
FCC-15B, Class B level (when shielded cables are used)
Slow - 6 readings per second
Medium - 48 readings per second (60 Hz)
Fast - 143 readings per second (20 configured channels)
Single Channel - 120 readings per second
Slow - 54 readings per second (60 Hz)
Medium - 200 readings per second
Fast - 1000 readings per second (20 configured channels)
Single Channel - 400 readings per second
Introduction and Specification
Characteristic
1
Specifications
1-11.
1-9