Sources Operation Manual Introduction Moxtek HPC 150W 75kV X-ray tubes are designed to be used in XRF and imaging applications. The tube is designed to be air-cooled and the cooling can be adapted to meet specific needs. The push/pull fan configuration makes it easy to direct the warm air outward.
Inspect the high voltage cable for proper tight and solid strain relief at the base of the tube. In the unlikely event that a tube has been damaged during shipping, note the damage and contact MOXTEK immediately. The serial number is...
#10 bolts can be used in the thru-holes to capture the assembly. Figure 4 - MoxHPC threaded hole mounts Cone Beam Configuration Description Angle Cone Angle 45° Center Line to 52.5° Central Ray Angle Figure 5 - Cone Beam Configuration MOXTEK TUB-MAN-1020 Rev C ® Subject to technical change without notice...
See figure 8 below for the circuit needed for the use of tachometer. Figure 8 - Circuit needed for tachometer MOXTEK TUB-MAN-1020 Rev C ® Subject to technical change without notice...
Page 5
5. Clean the cable connector with a lint free wipe. Figure 10 - Cleaning the cable connector 6. Apply a small amount of silicone oil and spread it around the connector. Figure 11 - Applying silicone oil MOXTEK TUB-MAN-1020 Rev C ® Subject to technical change without notice...
Page 6
NOTE: The filament monitor current of Spellman power supplies have been shown to have significant variation. If you find that the desired beam current cannot be attained, please contact Moxtek for guidance on using increased current limits. The X-ray tube may be permanently damaged in only a few seconds if the filament is run at excessive current levels.
Precautions should be taken to protect the operator while applying high voltages to avoid serious injury or death. 452 West 1260 North / Orem, UT 84057 Phone 801.225.0930 / Fax 801.221.1121 www.moxtek.com info@Moxtek.com For warranty and ordering information, please visit www.moxtek.com. MOXTEK TUB-MAN-1020 Rev C ® Subject to technical change without notice...
Need help?
Do you have a question about the X-ray Sources MOX-HPC150 Max and is the answer not in the manual?
Questions and answers