Pre-use Inspection

The Laparoscopic Ultrasound probe must be inspected prior to every
exam.
- By running your fingers about the exterior surfaces of the
transducer, articulating section, and rigid shaft, carefully
inspect for any signs of mechanical damage such as cracks, cuts,
tears, perforations, or protrusions. Any such damage could
compromise the electrical safety of the probe or cause lacerations
to the patient.
- Inspect the probe cable for damage, i.e. nicks, cuts, or tears in
the cable jacket.
- Check the control levers for proper operation. The levers should
move freely, without binding, and easily move the transducer
through its full range of movement.
If any damage is detected, do not use the probe. There are no
user serviceable parts associated with this instrument. Return
the probe to us or our authorized agent for service.