Hitachi EUP-OL334 Instruction Manual page 9

Laparoscopic probe
Table of Contents

Advertisement

1)
Transducer
2)
Articulating section
3)
Inserting section
4)
Grip section
5)
Left lever
6)
Right lever
7)
Cable
8)
System end connector
9)
Watertight action
10) Non watertight action : Section that should not be poured on section
Never use the probe which transducer and inserting section show
any damage such as cracks or cuts, or which cable jacket is broken.
If using as it is damaged, electric shock may result.
Connector section of the probe is not watertight. Never pour any
liquid on it or immerse it in any liquid. If neglected, it may
be damaged.
: Transmits and receives ultrasound.
: Section to articulate the transducer tip in
the up/down plane or right/left plane with
the right or left lever.
: Section that can be inserted into the trocar.
: Section to grip the probe.
: Lever to control articulation in the up/down
plane.
: Lever to control articulation in the
right/left plane.
: Cable to connect the grip and the system end
connector.
: Section to connect this instrument with the
ultrasound scanner unit.
: Section that can be immersed in the sterilant
solution and section cleaning solution.
with water or the section sterilant solution,
or should not be immersed in it. Otherwise
it may be damaged. Never immerse it in water
or the sterilant solution.
WARNING
CAUTION