Hitachi EUP-OL334 Instruction Manual page 23

Laparoscopic probe
Table of Contents

Advertisement

Connector section of the probe is not watertight. Never pour
any liquid to it nor immerse it in any liquid. If neglected,
the probe may be damaged.
The following should be observed before connecting/
disconnecting the probe. If doing so in the state other than the
following, malfunction or fault of the equipment may result.
Avoid such operation.
• Turn the power switch OFF.
• For any model having more than two probe connectors, select
the other connector than that for the probe to be
connected/disconnected using the probe selection key to make
Freeze ON.
Great care must be exercised to handle the probe.
Never give shock, hit any hardware, nor drop it. Otherwise, fault
may result.
Do not use 5mm/10mm, 5mm/11mm and 5mm/12mm trocars. Using these
may damage the probe.
Ultrasound jelly supplemented to the ultrasound scanner unit is
not sterilized. Never use it if the probe is to be used during
operation.
CAUTION
REMINDER
CAUTION
CAUTION
CAUTION
-17-
Q1E-EP0304