Hitachi EUP-OL334 Instruction Manual page 15

Laparoscopic probe
Table of Contents

Advertisement

Point of use (Pre-cleaning)
6.1
Pre-cleaning should be done immediately after each
use. The procedure is as follows:
1) Remove the protective cover.
2) Clean the probe of all patient's blood or fluid with running tap water
until the surface of the probe looks visually clean.
3) Wipe the whole surface of the probe with gauze pad and remove superficial
visible impurities.
Containment and transportation
6.2
Putting the contaminated equipment into exclusive
shock and damage proof container for transportation is recommended. It
is recommended that instruments are reprocessed as soon as possible and
not later than 4 hours after usage.
Manual Cleaning and disinfection
6.3
Prepare
following
cleaning and disinfection:
a) Detergent: Cidezyme (Johnson & Johnson,
#2258)or another cleaning agent with
approved material compatibility for this
medical device.
b) Disinfectant: Cidex OPA (Johnson & Johnson,
# 20391)or another disinfectant with
approved material compatibility for this
medical device.
c) Two tanks, one for cleaning and one for
disinfection - optional:
1 additional tank for rinsing with deionized/tap water (sufficient size
for immersion of the immersible part of the probe at full length)
d) Soft, fluff free cloth or single use towel
e) Personal protective equipment (gloves, water repellent protective
items
before
-9-
Manual cleaning and
manual
disinfection
Manual Cleaning
Rinsing after manual
Manual Disinfection
Rinsing after manual
disinfection
Point of use
(Pre-cleaning)
Containment and
transportation
cleaning
Q1E-EP0304