Hitachi EUP-OL334 Instruction Manual page 16

Laparoscopic probe
Table of Contents

Advertisement

skirt, face protection mask or protective glasses, see also
instructions of the manufacturer for the detergent and the
disinfectant)
Manual Cleaning:
Prepare the detergent solution in a tank with cold water (please follow
the instructions of the detergent manufacturer regarding application,
dilution and contact time).
1)
The temperature of the detergent solution should be between 15-30 °C,
concentration is 1.6%. Please note the minimum contact time of the
detergent in the manufacturer's instruction. If a differing detergent
is used, please also note the approved material compatibility for the
medical device.
2)
Immerse the immersible part of the probe without connector into the
diluted detergent solution (see Figure 3). Wipe the immersible part
of the probe under the surface of the detergent solution with a soft
cloth to remove all visible soil. Be sure that all grooves of the probe
are implemented during the cleaning process.
3)
The immersible part of the probe should be left in the detergent
solution according to the specified contact time of the detergent
manufacturer.
4)
Wipe the un-immersible parts of the probe with a soft cloth dipped
with the detergent solution.
5)
Rinse the immersible part of the probe with running tap water for 1
minute. (alternatively: immerse the immersible part of the probe in
a tray filled with deionized water/tap water (see Fig.3) for 5 min.)
6)
Visually check the outer surface of the probe for cleanness. If
necessary, use magnifying glass for visually check. If there is still
soil visible, repeat all above steps.
Manual disinfection:
1) Prepare the disinfectant solution in a tank with cold water (please
follow the instructions of the disinfectant manufacturer regarding
application, concentration, microbiological efficiency, service life
-10-
Q1E-EP0304