Artifact in DR Detector images

Details
Cause
Brief Solution
An artifact is visible in the images produced by a DR
Detector.
Exposure conditions have changed significantly since
latest calibration.
Perform calibration of the DR Detector.
For details, refer to the DX-D DR Detector Calibration
Key User Manual (0134).
DR 14s | Problem solving | 99