Cal Step
Input ABCD Gain Calibration Points (see
CL 0799
5 mV, 50 Hz
CL 0800
12.5 mV, 50 Hz
CL 0801
25 mV, 50 Hz
CL 0802
50 mV, 50 Hz
CL 0803
125 mV, 50 Hz
CL 0804
250 mV, 50 Hz
CL 0805
500 mV, 50 Hz
CL 0806
1.25 V, 50 Hz
CL 0807
2.5 V, 50 Hz
CL 0808
5 V, 50 Hz
CL 0809
12.5 V, 50 Hz
CL 0810
25 V, 50 Hz
CL 0811
50 V (set 5502A to OPR!), 50 Hz
CL 0812
125 V, 50 Hz
CL 0813
250 V, 50 Hz
DMM Meter Zero Calibration Points (2-channel models only)
CL 0890
no input, see
DMM Volt Gain Calibration Points (see
CL 0840
500 mV, 0 Hz
CL 0849
2.5 V, 0 Hz
CL 0841
5 V, 0 Hz
CL 0842
50 V (set 5502A to OPR!), 0 Hz
CL 0843
500 V, 0 Hz
CL 0844
1000 V, 0 Hz
Ohm Gain Calibration Points (see
CL 0910
100 Ω
CL 0911
1 kΩ
CL 0912
10 kΩ
CL 0913
100 kΩ
CL 0914
1 MΩ
CL 0915
10 MΩ
CL 0916
100%: READY
Table 14. Calibration Points (cont.)
DUT Input Signal
Figure
Multimeter Meter Zero
Figure
Figure
20)
18)
Normal output
SCOPE Off, 0 V, 0 Hz, Operate
19)
SCOPE Off, 500 mV, 0 Hz
SCOPE Off, 2.5 V, 0 Hz
SCOPE Off, 5 V, 0 Hz
SCOPE Off, 50 V, 0 Hz, Operate
SCOPE Off, 500 V, 0 Hz
SCOPE Off, 1000 V, 0 Hz
SCOPE Off, 100 Ω, Operate
ScopeMeter® Test Tool
Calibration Adjustment
5502A Setting
53