Disassembly and Reassembly Procedures

This section provides the required disassembly procedures. The printed circuit assembly
removed from the Test Tool must be adequately protected against damage.
The Test Tool contains static sensitive components. Handling and servicing these
components should be done only at a static free workstation by qualified personnel.
The Test Tool contains a Li-ion battery pack. See Section 1 for instructions on how to safely
handle and use this battery pack. The Users Manual is available at www.fluke.com.
The Test Tool uses self-tapping screws. For longer life, use a hand-operated screwdriver to
reinsert the screws into the same screw-hole threads on the case.
See
Figure
22,
Figure
To prevent electric shock, disconnect test leads, probes and power supply from
any live source and from the Test Tool itself. Always remove the battery pack
before completely disassembling the Test Tool. Only qualified personnel using
customary precautions against electric shock should work on a disassembled unit
with power on.

Required Tools

To access all the assemblies, you need:
Static-free work surface and anti-static wrist wrap
#10 Torx screwdriver
A small screwdriver or pair of tweezers to unlock flat cables from their connector
Cotton gloves (to avoid contaminating the lens and the PCA)

Remove the Tilt Stand, Hang Strap, and Side Strap

To separate the tilt stand from the rear case:
1. Gently bend one rotation point away from the rear case.
2. Move the stand away from the housing.
There is no need to remove screws or other fixing devices.
3. Before opening the Test Tool, remove the hang strap and the side strap. See the Users
Manual.
The grip of the side strap consists of two halves kept together with hook-and-loop tape.
The straps can be taken apart and removed from their fixing dowels in the side of the Test
Tool. Before you do this, take careful notice on the correct position of the strap.
4. To install, work in reverse order.
23, and
Figure 24
for disassembly.
XW Warning
ScopeMeter® Test Tool
Disassembly and Reassembly Procedures
55