Input D Lf-Hf Gain - Fluke ScopeMeter 190 III Series Service Manual

Hide thumbs Also See for ScopeMeter 190 III Series:
Table of Contents

Advertisement

Wait until the display shows calibration status CL 0549: READY.
2
5. Press
point signal, and start the calibration.
Continue through all calibration points of Input C LF-HF Gain.
6. Wait until the display shows calibration status CL 0465: READY.
7. When you are finished, set the 5502A to Standby.
8. Continue at the

Input D LF-HF Gain

To do the Input D LF-HF Gain calibration:
2
1. Press
2. Connect Ch. D of the Test Tool to the 5502A. See
3. Set the 5502A SCOPE output to source the signal required for the first calibration point in
Input D LF-HF Gain (CL 0675, Pos D Fast).
4. Set the 5502A to operate (OPR) or standby (STBY) as indicated.
3
5. Press
Wait until the display shows calibration status CL 0569: READY.
2
6. Press
point signal, and start the calibration.
Continue through all calibration points of Input D LF-HF Gain.
7. Wait until the display shows calibration status CL 0467: READY.
8. When you are finished, set the 5502A to Standby.
9. Continue at the
to select the next calibration step, set the 5502A to the next calibration
Input D LF-HF Gain
section.
to select the first calibration step in
to start the calibration.
to select the next calibration step, set the 5502A to the next calibration
Input AB Position (All Models)
Table
14, Input D LF-HF Gain.
Figure
16.
section.
ScopeMeter® Test Tool
Calibration Adjustment
43