Specifications; Product Specifications - YOKOGAWA 701933 User Manual

Current probe
Table of Contents

Advertisement

Specifications

Product Specifications

IM 701933-01E
Accuracy of 23 ±3 °C, 30 minutes after turning ON the power.
Frequency band*
Rise time*
Maximum continuous input
Maximum peak current 50 Apeak, non-continuous
Output voltage rate*
Amplitude accuracy*
Noise*
Insertion impedance
Temperature coefficient for sensitivity*
Propagation delay
Maximum rated power 5.6 VA (within maximum input)
Rated supply voltage
Operating temperature and humidity
Storage temperature and humidity
Operating location
Effect of external magnetic fields
Measurable conductors Insulated conductors
Diameter of measurable conductors
Guaranteed accuracy period
Cable lengths
External dimensions
Weight
Accessories
* In conjunction with a waveform measuring instrument having an input impedance of 1
MΩ ±1%
DC to 50 MHz (–3 dB)
(see the typical characteristics shown on the next page, figure 1)
7 ns or less
30 Arms (AC and DC components)
(See the next page, figure 2 for derating according to frequency)
0.1 V/A
0 to 30 Arms: ±1.0% of reading ±1 mV
30 Arms to 50 Apeak: ±2.0% of reading
(DC, 45 to 66 Hz)
Equivalent to 2.5 mArms or less
(for 20 MHz band measuring instrument)
(See typical characteristics on the next page, figure 3)
± 2% (with 50 Hz, 30 Arms input in a range of 0 to 40 °C)
13 ns (Typical)
±12 V ±0.5 V
0 to 40 °C, 80% RH or less (no condensation)
–10 to 50 °C, 80% RH or less (no condensation)
Indoor use, altitude up to 2000 m
Equivalent to a maximum of 20 mA
(in a DC or 60 Hz, 400 A/m magnetic field)
φ 5 mm
1 year (10000 open and close operations)
Sensor cable: approx. 1.5 m
Power supply cable: approx. 1 m
Sensor: approx. 175 (W) × 18 (H) × 40 (D) mm
Terminator: approx. 27 (W) × 55 (H) × 18 (D) mm
Approx. 230 g
User's manual, soft case
17